This paper presents research dynamic properties of a calibration comparator which is used to calibrate high precision line standards of length. For this purpose, multi-body dynamic and mathematical models of a carriage system of the comparator were presented. Calculated amplitude-frequency responses and modes of oscillations allowed a determination of resonant frequencies of the system.