وثيقة
Severe intraoperative hypercapnia complicating an unsual malfunction of the inner tube of a co-axial (BAIN'S) circuit.
المساهمون
Iyasere, Godwill., مؤلف
الناشر
Oman Medical Specialty Board.
ميلادي
2010-04
اللغة
الأنجليزية
الملخص الإنجليزي
The Bain's co-axial circuit system is fully established in general anaesthesia practice. It is favoured for its light weight and suitability for head and neck surgery. However, there are numerous published reports of malfunction of the inner tube of the Bain's co-axial circuit, with potentially lethal complications for the patient. This report presents a case in which a patient connected to a reused Bain's circuit (Datex-Ohmeda) developed severe hypercapnia in the early intraoperative period due to unusual defect of the inner tube. This report tests and outlines the integrity of co-axial circuits and also reviews the available literature.
المجموعة
URL المصدر
zcustom_txt_2
Youssef, Youssef Emam, & Iyasere, Godwill (2010). Severe intraoperative hypercapnia complicating an unsual malfunction of the inner tube of a co-axial (BAIN'S) circuit. Oman Medical Journal, 25 (2).
قالب العنصر
مقالات الدوريات