Document

Severe intraoperative hypercapnia complicating an unsual malfunction of the inner tube of a co-axial (BAIN'S) circuit.

Contributors
Publisher
Oman Medical Specialty Board.
Gregorian
2010-04
Language
English
English abstract
The Bain's co-axial circuit system is fully established in general anaesthesia practice. It is favoured for its light weight and suitability for head and neck surgery. However, there are numerous published reports of malfunction of the inner tube of the Bain's co-axial circuit, with potentially lethal complications for the patient. This report presents a case in which a patient connected to a reused Bain's circuit (Datex-Ohmeda) developed severe hypercapnia in the early intraoperative period due to unusual defect of the inner tube. This report tests and outlines the integrity of co-axial circuits and also reviews the available literature.
Member of
Citation
Youssef, Youssef Emam, & Iyasere, Godwill (2010). Severe intraoperative hypercapnia complicating an unsual malfunction of the inner tube of a co-axial (BAIN'S) circuit. Oman Medical Journal, 25 (2).
Category
Journal articles